|
|
|
 |
 |
 |
|
 |
Recirculating chillers can be ordered from Lytron or from one of our distributors. Order directly from Lytron by calling +1-781-933-7300. Standard systems can also be ordered from one of our international distributors. | First Select Your Model Number | | | | RC006 | RC009 | RC011 | RC022 | RC030 | RC045 | | Cooling capacity1 | W
kBTU/Hr | 825
2.8 | 1050
3.6 | 1650
5.6 | 2400
8.2 | 3450
11.8 | 5900
20.1 | | Compressor capacity | HP | 1⁄4 | 1⁄3 | 1⁄2 | 3⁄4 | 1 | 1 1⁄2 | | Temperature stability2 | | ± 0.1°C | | Fluid connections | | 1⁄2˝ FNPT | | Reservoir capacity | | 1 gal / 4 liters | 2 gal / 8 liters | 6 gal / 23 liters | | Coolant temperature range | | 40°F to 95°F / 5°C to 35°C | | Ambient temperature range | | 50°F to 95°F / 10°C to 35°C | | Dimensions (W x D x H) | inches
mm | 12.5 x 19.0 x 22.0
318 x 483 x 559 | 14.8 x 24.5 x 26.5
376 x 623 x 673 | 21.4 x 27.8 x 31.9
543 x 705 x 810 | | Weight | dry lbs (kg) | 97 (44) | 100 (45) | 122 (55) | 166 (75) | 260 (118) | 270 (122) | | Next, Pick An Electrical Configuration | | Electrical configurations and full load amperage³ | | G03: 115V, 60 Hz, 1ph | Amps | 9.9 | 12.2 | 14.3 | n/a | n/a | n/a | | H03 : 230V, 50 Hz, 1ph | Amps | 4.5 | 5.3 | 6.3 | 9.5 | 13.7 | 17.2 | | J03: 208/230V, 60 Hz, 1ph | Amps | n/a | 5.8 | 7.4 | 10.0 | 14.5 | 19.6 | | Now, Select A Pump | | Pump options (refer to selecting a pump and system pump graphs) | | BB: PDP4, Brass, 1.3 gpm / 4.9 lpm | x | x | x | x | | | | BC: PDP4, Brass, 1.8 gpm / 6.8 lpm | o | o | o | o | | | | BE: PDP4, Brass, 2.3 gpm / 8.7 lpm | o | o | o | o | | | | BG: PDP4, Brass, 4.3 gpm / 16.3 lpm | | | o | o | x | x | | CB: PDP4,5, Stainless Steel, 1.3 gpm / 4.9 lpm | o | o | o | o | | | | CC: PDP4,5, Stainless Steel, 1.8 gpm / 6.8 lpm | o | o | o | o | | | | CE: PDP4,5, Stainless Steel, 2.3 gpm / 8.7 lpm | o | o | o | o | | | | CG: PDP4,5, Stainless Steel, 4.3 gpm / 16.3 lpm | | | o | o | o | o | | AA: Centrifugal, 1⁄20 HP6 | o | o | | | | | | DA: Centrifugal, 1⁄4 HP6 | | | o | o | o | o | | DD: Centrifugal, 1⁄2 HP6 | | | o | o | o | o | | EC: Turbine, 1⁄6 HP6 | o | o | o | o | | | | EB: Turbine, 1⁄2 HP6 | | | o | o | o | o | | FB: Stainless Turbine, 1⁄2 HP6 | | | o | o | o | o | | Select Your Controller | | Controller options | | Package 17: Digital temperature display, calibration offset, low flow shut-off, auto-restart, °C/ °F toggle. | o | o | o | o | o | o | | Package 2: Package 1 plus digital pressure sensing, low level, low/high temperature, pressure display, fault shut-off (toggle on/off), relay contacts, audible alarm and alarm mute. | x | x | x | x | x | x | | Package 3: Package 2 plus RS232. | o | o | o | o | o | o | | Add Any Additional Options | | Available options (full description of these options can be found below) | | External flow valve | o | o | o | o | o | o | | External pressure relief valve | o | o | o | o | o | o | | Anti-siphon system | o | o | o | o | o | o | | Air filter | o | o | o | o | o | o | | 5 micron coolant filter8, 10 | o | o | o | o | o | o | | Heater9 | | | o | o | o | o | | Internal insulation package | o | o | o | o | o | o | | Low temperature operation | | o | o | o | o | o | | Water-cooled condenser | | | o | o | o | o | | 0.1°C set point | o | o | o | o | o | o | | DI water cartridge8 | o | o | o | o | o | o | | High purity plumbing | o | o | o | o | o | o | | PAO compatibility | | | o 9 | o | o | o | | Remote start | o | o | o | o | o | o | | To Arrive At A Part Number | RC011 G03 BB 2 M__ = RC011 chiller with G03 (115V, 60 Hz, 1ph) electrical configuration, a BB pump, and controller package 2 Customization options (A 4 digit option code will be assigned at time of order, based on selected options. Leave blank if no additional options selected) x = standard o = available option | | | 1At 20°C setpoint, 20°C ambient, 60Hz input supply 2Assumes stable load 3With standard pump 4PDP = Positive Displacement Pump 5Only available with high purity plumbing 6Actual flow rate depends on system pressure drop. 75 piece min order 8Not available with AA and DA pump 9Not available in G03 electrical configuration 10Recommended when selecting high purity option |
Cooling System Kodiak Options| Option | Benefit | | Protection Options | | External flow valve: HIGHLY RECOMMENDED. A manually adjustable globe valve installed on the back of the chiller in parallel with the chiller coolant supply/return. | Allows the operator to control the flow rate to the application. As the external flow valve is opened, more of the flow bypasses the application. | | External pressure relief valve: A manually adjustable pressure relief valve (50 to 100 psi or 0 to 50 psi) installed on the back of the chiller in parallel with the chiller coolant supply/return. | Allows the chiller to continue to run and maintain temperature if the flow to the application is interrupted. When flow to the application is stopped, the pressure relief valve opens and bypasses flow to the chiller’s return. It can also be used to prevent the chiller supply from exceeding a predetermined pressure setting. | | Anti-siphon system: Check and solenoid valves installed internal to the chiller on the supply and return lines, respectively. | Allows the chiller to be installed at an elevation below the application. Prevents backflow and subsequent overflowing of the chiller’s vented tank (if not installed, fitting leaks in the application can allow air to displace the coolant, resulting in backflow to the tank). | | Air filter: Internally mounted on the air inlet to the condenser coil. | Filters incoming air to the chiller’s condensing coil, preventing dust build up which can lead to decreased performance. The easy-access filter is simple to replace, reducing the need to clean the condensing coil and refrigeration components. | | 5 micron coolant filter: Externally mounted on the supply from the chiller. | Filters process coolant being supplied to the application, protecting equipment from blockage or damage due to particulate buildup and extending pump life. | | Special Applications Options | | Heater: Submerged 2,000 Watt electric resistance heater with built-in over temperature shut-off (non-adjustable). | Allows chiller to reach elevated set points faster (useful for applications in cold start-up situations or where the coolant set point is frequently cycled). This feature does not extend the coolant temperature range of the chiller. | | Internal insulation package: Insulated tank, pump head, and coolant lines. | Eliminates condensation forming on the components in contact with the fluid when operating chiller at set points below the ambient dew point. | | Low temperature operation: Chiller coolant temperature range is –5°C to +25°C. (Note: internal insulation package must also be selected). | Optimizes performance for low temperature operation and provides additional low temperature cooling capacity compared to standard chillers. Contact Lytron’s applications engineering department for performance graphs. | | Water-cooled condenser: Refrigerant is condensed using water (rather than ambient air) via a liquid-to-liquid heat exchanger. | Heat that would normally be rejected to the ambient environment is transferred to an external liquid coolant source, avoiding room warming and possible air conditioning overload. Also, water-cooled chillers are quieter because there is no condenser fan. Contact Lytron for required facility liquid coolant flow rate and temperature requirements. | | 0.1°C set point: Allows 0.1°C increments of the set point and display temperatures. | Allows more precise control of the coolant delivery temperature. Available in °C only. | | Fluid Compatibility Options | | Deionization (DI) water cartridge: Externally mounted DI cartridge (no indicator light) on the return to the chiller. | Filters ions from coolant to maintain a fluid resistivity level between 1 and 3 megohm/cm by partial flow through the resin bed. This feature does not include a resistivity indicator light. High purity plumbing is recommended with the DI water cartridge. | | High purity plumbing: Includes nickel-brazed evaporator and nickel-plated or nylon fittings compatible with DI water. Not compatible with brass pumps. | All wetted materials are fully compatible with DI water to prevent component corrosion. | | PAO compatibility: Includes pump, fittings, hose, and oil filter compatible with polyalphaolefin (PAO). Anti-siphon system and internal insulation packages are included in this option. | All wetted materials are fully compatible with PAO. Contact Lytron for PAO performance curves. | | Convenience Options | | Remote start: Chiller power can be cycled through a computer or relay. | Allows an external circuit to control the chiller on/off function via dry contacts on the rear of the chiller. | | Please refer to the chiller technical manual for further discussion of these features. |
Cooling System Kodiak Controllers| Controller Option | Functionality | Controller Package | | | 1 | 2 | 3 | | Digital temperature display | Easy-to-read display in either °C or °F. | • | • | • | | Calibration offset | Allows temperature display to be offset to represent the temperature at a different point within the cooling circuit. | • | • | • | | Low flow shut-off fault | Alerts user to a lack of coolant flow and shuts down the chiller to prevent the refrigeration system from freezing. | • | • | • | | Auto-restart | Automatically returns chiller to its operating status prior to interruption of power, e.g., if the chiller was on, it will turn back on when power is restored. | • | • | • | | °C/°F toggle | Enables temperature display to be viewed in °C or °F. | • | • | • | | Audible alarm | Audible warning of all fault conditions. | • | • | • | | Alarm mute | Silences audible alarm until the alarm is cleared or the chiller turned off. | • | • | • | | Digital pressure sensing | Displays pump output pressure (psig or bar) and allows it to be transmitted through RS232. | | • | • | | Low level fault | Alerts user that the coolant level in the reservoir is at a critically low level (approaching the level of the pump’s suction line). Switches chiller off if ‘fault shut-off’ is on. | | • | • | | Low/high temperature fault | Alerts user that the coolant temperature is outside of the user’s defined set-point limits. Switches chiller off if ‘fault shut-off’ is on. | | • | • | Fault shut-off (toggle on/off) | Allows chiller to be programmed to shut down on fault (on) or continue to run during fault conditions (off). The chiller will always shut down on the low flow shut-off alarm. | | • | • | Relay contacts (normally open or normally closed) | Fault conditions of the chiller can be monitored by applying a low voltage power signal across dry contacts at the rear of the chiller. (Power loss to the chiller is not a fault condition.)* Mating connector—AMP part number 350715-1 and 350547-1 | | • | • | | RS232 communications | Allows complete control of all the chiller’s controller features (changing/monitoring coolant set point, monitoring pressure, and fault alarms, etc.) from a remote computer for monitoring, control, and data logging. | | | • |
Package 2 is the standard package.
Cooling System Kodiak Accessories| Item | Description | Part Number | | Water Filter Kit: | 1⁄2˝ fluid connections. Contains 5 micron filter, filter housing, mounting bracket ,1´ hosing, base clamps, and assembly instructions. | 200-0441 | | Plumbing package: | 1⁄2˝ fluid connections. Contains 50’ Nylabraid™ hose, 50’ Armorflex™ insulation, male NPT to barb fittings, hose clamps. | 200-0193 | External pressure relief valve kit:
(50 - 100 psi) | 1⁄2˝ fluid connections. For applications requiring reduced pressure. This kit contains a pressure relief valve, 2 tees, hose clamps, and plumbing instructions. Assembly required. | 200-0196 | External pressure relief valve kit:
(0 - 50 psi) | 1⁄2˝ fluid connections. For applications requiring reduced pressure. This kit contains a pressure relief valve, 2 tees, hose clamps, and plumbing instructions. Assembly required. | 205-0080 | | External manually adjustable flow valve: | 1⁄2˝ fluid connections. For applications requiring flow control. Kit contains flow valve, 2 tees, hose clamps and plumbing instructions. Assembly required. | 200-0298 | | 115V power cord: | 25’ | 250-0092 | | 230V power cord: | 25’ | 250-0093 | | RS232 cable: | 10’ | 250-0091 | | Dry contact cable: | 10’ | 250-0159 | | 5 micron replacement coolant filters* | 330-0022 | | Air filters* | RC006/RC009: 330-1541
RC011/RC022: 330-0977
RC030/RC045: 330-0683 | | Deionization cartridge | 430-0330 |
*10 piece minimum
|
|
Complete our short form to request a quote or catalog or ask an engineer a technical question.
|
 |
 |
 |
|